FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3820289 · Received May 20, 2014

Report

Report Number
3004209178-2014-84827
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
February 27, 2014
Report Date
May 9, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE DEVICE WAS RECEIVED WITH SCRATCHED SCREEN, CRACKED CASE DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER REPORTED HAVING ISSUES WITH THE INSULIN PUMP. THE DEVICE ALARMED WHILE ATTEMPTING TO CHANGE THE INFUSION SET. THE BLOOD GLUCOSE READING WAS 120MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300477 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 57 YR