FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3820278
·
Received May 20, 2014
Report
- Report Number
- 3004209178-2014-84803
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- February 18, 2014
- Report Date
- May 7, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LO
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD SCRATCHED SCREEN, CRACKED DISPLAY WINDOW CORNERS, AND CRACKED RESERVOIR TUBE LIP. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED. THE DRIVE SUPPORT CAP DID NOT APPEAR TO BE DAMAGED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 299490 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754WWP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |