FDA Adverse Event Malfunction Summary report: N

MAESTRO MEDIUM FIXED DURAGUARD

MDR report key: 3820023 · Received May 20, 2014

Report

Report Number
0001811755-2014-01835
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 22, 2014
Report Date
April 22, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING FAILURE ANALYSIS, THE REPORTED EVENT OF A BENT DURAGUARD FOOT WAS CONFIRMED BY THE TECHNICIAN THROUGH VISUAL EVALUATION. A BENT DURAGUARD CAN OCCUR IF EXCESSIVE SIDE LOAD WAS APPLIED TO THE FEATURE. THIS IS NOT A REPAIRABLE DEVICE; THEREFORE, IT WAS NOT RETURNED TO THE USER FACILITY FOLLOWING EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE FOOT OF THE MAESTRO MEDIUM FIXED DURAGUARD WAS BENT DURING SERVICE INSPECTION AT THE MANUFACTURING FACILITY. NO PATIENT INVOLVEMENT, NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300064 MAESTRO MEDIUM FIXED DURAGUARD MOTOR, DRILL, PNEUMATIC HBB STRYKER INSTRUMENTS-KALAMAZOO 12116

Patients

Seq Age Sex Outcome Treatment
1