FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION WITH TELEPACK

MDR report key: 3813549 · Received March 26, 2014

Report

Report Number
2221819-2014-00117
Event Type
Malfunction
Date Received
March 26, 2014
Date of Event
November 26, 2013
Report Date
November 27, 2013
Manufacturer
MINDRAY DS USA, INC
Product Code
MHX
PMA / PMN Number
031760
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REP EVALUATED THE SYSTEM. CORRECTIONS INCLUDED CLEARING THE SYSTEM'S ERROR LOGS AND REBOOTING. COMMUNICATION WAS REESTABLISHED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE PANORAMA CENTRAL STATION HAD LOST COMMUNICATION, WHICH MAY HAVE AFFECTED TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
176891 PANORAMA CENTRAL STATION WITH TELEPACK CENTRAL MONITORING SYSTEM MHX MINDRAY DS USA, INC

Patients

Seq Age Sex Outcome Treatment
1 UNK