FDA Adverse Event
Malfunction
Summary report: N
PANORAMA CENTRAL STATION WITH TELEPACK
MDR report key: 3813549
·
Received March 26, 2014
Report
- Report Number
- 2221819-2014-00117
- Event Type
- Malfunction
- Date Received
- March 26, 2014
- Date of Event
- November 26, 2013
- Report Date
- November 27, 2013
- Manufacturer
- MINDRAY DS USA, INC
- Product Code
- MHX
- PMA / PMN Number
- 031760
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
COMPANY REP EVALUATED THE SYSTEM. CORRECTIONS INCLUDED CLEARING THE SYSTEM'S ERROR LOGS AND REBOOTING. COMMUNICATION WAS REESTABLISHED.
Description of Event or Problem · 1
CUSTOMER REPORTED THE PANORAMA CENTRAL STATION HAD LOST COMMUNICATION, WHICH MAY HAVE AFFECTED TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 176891 | PANORAMA CENTRAL STATION WITH TELEPACK | CENTRAL MONITORING SYSTEM | MHX | MINDRAY DS USA, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |