FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3813464 · Received May 15, 2014

Report

Report Number
2531779-2014-13727
Event Type
Malfunction
Date Received
May 15, 2014
Report Date
September 16, 2025
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/06/2014 WITH THE FOLLOWING FINDINGS:A PORTION OF THE KEYPAD WAS FOUND TO BE MISSING. EVALUATION REVEALED THE DOWN KEYPAD BUTTON WAS INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS. THE DOWN BUTTON CONTACT WAS FOUND TO BE INVERTED. INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND DISCOLORED. A BATTERY COMPARTMENT CRACK WAS OBSERVED DURING EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DISPLAY ISSUE, KEYPAD DAMAGE AND RESPONSE ISSUE, AND CASE DAMAGE ISSUE. THIS REPORT IS MADE BASED ON RESULTS OF THE INVESTIGATION PERFORMED ON 05/06/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
292046 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 22 YR Female