PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG
Report
- Report Number
- 1419937-2014-00315
- Event Type
- Malfunction
- Date Received
- March 31, 2014
- Date of Event
- March 19, 2014
- Report Date
- March 26, 2014
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
A REPLACEMENT PUMP, INCLUDING TRANSFORMER, WAS SENT TO THE CUSTOMER. THIS ISSUE WITH THE DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORSEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING MONITORED FOR EFFECTIVENESS OF THE ABOVE MENTIONED CORRECTIVE ACTION.
THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT HER PUMP IN STYLE ADVANCED BREAST PUMP MAKES A CLICKING NOISE. UPON RECEIPT OF THE CUSTOMER'S TRANSFORMER AT MEDELA, IT WAS NOTED BY THE SERVICE CENTER THAT THE TRANSFORMER WAS BREACHED, WHICH IS A SAFETY RISK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 190016 | PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG | HGX | MEDELA, INC. | 57036/9207010 | REV M-1512 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |