FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG

MDR report key: 3813441 · Received March 31, 2014

Report

Report Number
1419937-2014-00315
Event Type
Malfunction
Date Received
March 31, 2014
Date of Event
March 19, 2014
Report Date
March 26, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT PUMP, INCLUDING TRANSFORMER, WAS SENT TO THE CUSTOMER. THIS ISSUE WITH THE DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORSEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING MONITORED FOR EFFECTIVENESS OF THE ABOVE MENTIONED CORRECTIVE ACTION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT HER PUMP IN STYLE ADVANCED BREAST PUMP MAKES A CLICKING NOISE. UPON RECEIPT OF THE CUSTOMER'S TRANSFORMER AT MEDELA, IT WAS NOTED BY THE SERVICE CENTER THAT THE TRANSFORMER WAS BREACHED, WHICH IS A SAFETY RISK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
190016 PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG HGX MEDELA, INC. 57036/9207010 REV M-1512

Patients

Seq Age Sex Outcome Treatment
1