FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3813366
·
Received May 15, 2014
Report
- Report Number
- 1416980-2014-15616
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Date of Event
- April 20, 2014
- Report Date
- April 20, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION, FUNCTIONAL TESTING, AND A REVIEW OF THE ALARM LOG WERE PERFORMED. VISUAL INSPECTION IDENTIFIED THE DAMAGE TO THE DOOR LATCH ROLLER. THE CAUSE OF THE DAMAGE TO THE DOOR LATCH ROLLER WAS NOT ABLE TO BE DETERMINED. TO CORRECT THE CONDITION, THE DOOR LATCH ROLLER WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED DOOR LATCH ROLLER. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290119 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |