HARMONIC ACE CURVED SHEARS INSERT
Report
- Report Number
- 2955842-2014-03003
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Date of Event
- April 10, 2014
- Report Date
- April 14, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K112584
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- OTHER
Narratives
THE INSERT HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE HAS NOT BEEN DETERMINED. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SUBMITTED. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO FOLLOWING CONCLUSION: THE JAWS OF THE HARMONIC ACE CURVED SHEARS INSERT BROKE OFF THE INSTRUMENT AND FELL INTO THE PATIENT.
THE INSERT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THE METALIC TIP OPPOSITE SIDE OF THE TEFLON PAD BROKEN AND 0.669 LONG PIECE WAS MISSING FROM THE HARMONIC INSERT. NO OTHER DAMAGE WAS FOUND. FAILURE ANALYSIS CONCLUDED THAT THE DAMAGE WAS LIKELY DUE TO MISHANDLING.
ON (B)(4) 2014, ISI ATTEMPTED TO SUBMIT THIS REPORT ON TIME WITHIN REPORTING TIMEFRAME; HOWEVER, THE ESG DOWNTIME THAT OCCURRED FROM (B)(4) 2014 3:43 PM EST THROUGH THURS (B)(4) 2014 3:00 AM EST PREVENTED ISI TO SUBMIT THIS REPORT ON TIME.
IT WAS REPORTED THAT DURING A DA VINCI SUPRACERVICAL HYSTERECTOMY PROCEDURE, THE JAWS OF THE HARMONIC ACE CURVED SHEARS INSERT BROKE OFF AND FELL INTO THE PATIENT. ACCORDING TO THE SURGEON, ALL BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. ON (B)(4) 2014, INTUITIVE SURGICAL, INC. CONTACTED THE CUSTOMER TO OBTAIN ADDITIONAL INFORMATION REGARDING REPORTED EVENT AND OBTAINED FOLLOWING INFORMATION: THE BROKEN PIECES FROM HARMONIC ACE CURVED SHEARS INSERT WERE RETRIEVED LAPAROSCOPICALLY. NO ADDITIONAL SURGICAL PROCEDURES AND NO POST-OPERATIVE TEST WERE REQUIRED TO RETRIEVE THE PIECES. THE SURGEON WAS SEALING THE CERVIX WHEN THE INSERT TIP BROKE OFF. THE INSTRUMENT WAS INSPECTED PRIOR TO USE AND NO PHYSICAL DAMAGE WAS FOUND. THE INSTRUMENT AND THE INSERT WERE IN USE FOR ABOUT 15 MINUTES. THERE WAS NO ALLEGATION OF PATIENT INJURY OR HARM.