FDA Adverse Event Malfunction Summary report: N

DHS®/DCS® LAG SCREW 12.7MM THREAD/95MM-STERILE

MDR report key: 3813225 · Received May 15, 2014

Report

Report Number
3009450863-2014-10007
Event Type
Malfunction
Date Received
May 15, 2014
Date of Event
April 10, 2014
Report Date
April 22, 2014
Manufacturer
SYNTHES BALSTHAL
Product Code
KTT
PMA / PMN Number
PK791619
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO NONCONFORMANCE REPORTS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS THE PRODUCT IS ENTERING THE COMPLAINT SYSTEM. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: A MANUFACTURING EVALUATION WAS PERFORMED. THE POSITIONING GROOVES ON THE SCREW HAVE BEEN WIDENED UP. SYNTHES (B)(4) MANUFACTURED P/N 280.950S LOT NO 8642187. THE PARTS WERE MADE TO SE_420941 REV 'B'. NO NONCONFORMANCE REPORTS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORDS SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLIANT CONDITION. SLOT DIMENSIONS: DAMAGED AND WIDENED UP. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED 2 DYNAMIC HIP SYSTEM (DHS) SCREW WOULD NOT FIT THROUGH THE PLATE. NO PATIENT HARM WAS REPORTED. THIS IS REPORT 1 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291314 DHS®/DCS® LAG SCREW 12.7MM THREAD/95MM-STERILE APPLIANCE,FIXATION,NAIL KTT SYNTHES BALSTHAL 8642187

Patients

Seq Age Sex Outcome Treatment
1