FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT? PLUS

MDR report key: 3813171 · Received May 15, 2014

Report

Report Number
2134265-2014-02581
Event Type
Malfunction
Date Received
May 15, 2014
Date of Event
February 17, 2014
Report Date
April 17, 2014
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P110010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION FOUND THAT THE HYPOTUBE WAS BROKEN 213MM DISTAL TO THE STRAIN RELIEF. THE HYPOTUBE WAS KINKED AT VARIOUS LOCATIONS ALONG ITS LENGTH. THIS TYPE OF DAMAGE IS CONSISTENT WITH EXCESSIVE FORCE BEING APPLIED TO THE DELIVERY SYSTEM. THE TIP WAS SLIGHTLY FLARED. THIS TYPE OF DAMAGE IS CONSISTENT WITH MEETING AN OBSTRUCTION DURING AN ATTEMPT TO CROSS A LESION. NO ISSUES WERE NOTED WITH THE CRIMPED STENT AND BALLOON. THE BALLOON WAS TIGHTLY WRAPPED AND DID NOT APPEAR TO HAVE BEEN SUBJECTED TO POSITIVE PRESSURE. NO OTHER ISSUES WERE IDENTIFIED DURING THE PRODUCT ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(4) 2014. IT WAS REPORTED THAT CROSSING DIFFICULTIES ENCOUNTERED. THE TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT CIRCUMFLEX (LCX) ARTERY. THE LESION WAS PRE-DILATED WITH A MAVERICK 2.0X20 CATHETER BALLOON. A 2.50X38MM PROMUS ELEMENT PLUS STENT WAS SELECTED TO TREAT THE LESION BUT IT FAILED TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS GOOD. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED SHAFT BREAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291311 PROMUS ELEMENT? PLUS STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493918438250 16171554

Patients

Seq Age Sex Outcome Treatment
1 60 YR CATHETER BALLOON: MAVERICK 2..0X20