FDA Adverse Event Injury Summary report: N

PLATE

MDR report key: 3813105 · Received May 15, 2014

Report

Report Number
2520274-2014-11316
Event Type
Injury
Date Received
May 15, 2014
Report Date
April 22, 2014
Manufacturer
DEPUY SYNTHES
Product Code
HRS
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICAL THERAPIST

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. OH, J.K., HWANG, J.H., LEE, S.J., KIM, J. II DYNAMIZATION OF LOCKED PLATING ON DISTAL FEMUR FRACTURE. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY (2011) 131:535-539. THIS REPORT IS FOR AN UNKNOWN SEVEN-HOLE LCP-DF PLATE. WITHOUT A LOT NUMBER, THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AFTER THE SUBSEQUENT REVIEW OF THE FOLLOWING JOURNAL ARTICLE: OH, J.K., HWANG, J.H., LEE, S.J., KIM, J. II DYNAMIZATION OF LOCKED PLATING ON DISTAL FEMUR FRACTURE. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY (2011) 131:535-539. THE ARTICLE REPORTS A (B)(6) FEMALE WITH OSTEOARTHRITIS EXPERIENCED A NONUNION AFTER MIPO (MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS) FOR A LONG OBLIQUE FRACTURE OF THE DISTAL FEMUR. A SEVEN-HOLE LCP-DF PLATE (SYNTHES) WAS USED. A CORTICAL 4.5 MM SCREW WAS PLACED ACROSS FRACTURE SITE AS A REDUCTION SCREW TO REDUCE THE DISTAL FRAGMENT THROUGH THE DYNAMIC COMPRESSION UNIT OF A COMBINATION HOLE OF THE LOCKING COMPRESSION PLATE-DISTAL FEMUR. AT THIS TIME, THE NEAR CORTEX WAS OVER-DRILLED WITH A 4.5 MM DRILL BIT. POSTOPERATIVE RADIOGRAPHS SHOWED GOOD ALIGNMENT EXCEPT SMALL DEGREE OF ANTERIOR ANGULATION ON THE LATERAL VIEW. NO POST-OPERATIVE COMPLICATIONS WERE REPORTED. AFTER SIX MONTHS, RADIOGRAPHS FAILED TO SHOW ANY CALLUS FORMATION AND COMPUTED TOMOGRAPHY (CT) SCANS REVEALED A FRACTURE GAP. AUTHOR REPORTS THE CORTICAL SCREW ACTED LIKE A LAG SCREW AND MADE THE CONSTRUCT TOO STIFF WHICH CAUSED THE NONUNION. LAG SCREW WAS REMOVED TO ALLOW MOTION AT THE FRACTURE SITE. SIX MONTHS LATER RADIOGRAPHS SHOWED SOLID FRACTURE HEALING WITH BRIDGING CALLUS FORMATION ALONG THE MEDICAL CORTEX. THIS ARTICLE SUMMARIZES POSTOPERATIVE FINDINGS FROM OTHER PUBLISHED ARTICLES. THIS IS REPORT 2 OF 2 FOR COMPLAINT - (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290875 PLATE HRS DEPUY SYNTHES

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention