FDA Adverse Event Malfunction Summary report: N

TPS HANDPIECE CORD

MDR report key: 3813015 · Received May 15, 2014

Report

Report Number
0001811755-2014-01787
Event Type
Malfunction
Date Received
May 15, 2014
Date of Event
April 25, 2014
Report Date
April 25, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K943540
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON VISUAL INSPECTION, THE DEVICE WAS FOUND TO HAVE A WORN WEDGE. THE DEVICE WAS DISCARDED BY THE MANUFACTURER.

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED AFTER THE QUALITY INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING ROUTINE MAINTENANCE CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY,THE TPS HANDPIECE CORD CAUSED THE HANDPIECE TO RUN WITHOUT USER ACTIVATION. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING ROUTINE MAINTENANCE CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY THE TPS HANDPIECE CORD CAUSED THE HANDPIECE TO RUN WITHOUT USER ACTIVATION. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291194 TPS HANDPIECE CORD DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO 11206

Patients

Seq Age Sex Outcome Treatment
1