LIFEVEST WCD 4000 SYS
Report
- Report Number
- 3008642652-2014-01246
- Event Type
- Malfunction
- Date Received
- April 25, 2014
- Date of Event
- March 23, 2014
- Report Date
- April 21, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK THERAPY PAD MESSAGES, ADJUST BELT OR CHECK BELT MESSAGES) HAS BEEN CONFIRMED. UPON EVALUATION, THE ELECTRODE BELT FAILED INCOMING FUNCTIONALITY TESTING. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN WHITE PULSE WIRE IN THE TRUNK CABLE. THERE WAS NO SIGN OF EXTERNAL DAMAGE TO THE CABLE. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE OPEN WIRE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
A (B)(6) FEMALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT SHE WAS RECEIVING CHECK THERAPY PAD AND ADJUST BELT OR CHECK BELT MESSAGES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 251124 | LIFEVEST WCD 4000 SYS | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |