FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3812924
·
Received May 15, 2014
Report
- Report Number
- 2032227-2014-02364
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Date of Event
- April 24, 2014
- Report Date
- April 24, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP FAILED THE DISPLACEMENT TEST WITH UNITS REMAINING ON THE STATUS SCREEN. UNIT HAD MOTOR ERROR ALARMS DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNIT HAD CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS, SCRATCHED DISPLAY WINDOW.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT SHE HAD HIGH BLOOD GLUCOSE BECAUSE HER INSULIN PUMP WAS MALFUNCTIONING. CUSTOMER STATED THAT SHE HAD AN MRI WITH THE INSULIN PUMP ATTACHED TO HER. CUSTOMER STATED THAT THE INSULIN PUMP IS ALARMING MOTOR ERROR. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290332 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR |