FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3812924 · Received May 15, 2014

Report

Report Number
2032227-2014-02364
Event Type
Malfunction
Date Received
May 15, 2014
Date of Event
April 24, 2014
Report Date
April 24, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT TEST WITH UNITS REMAINING ON THE STATUS SCREEN. UNIT HAD MOTOR ERROR ALARMS DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNIT HAD CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS, SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE HAD HIGH BLOOD GLUCOSE BECAUSE HER INSULIN PUMP WAS MALFUNCTIONING. CUSTOMER STATED THAT SHE HAD AN MRI WITH THE INSULIN PUMP ATTACHED TO HER. CUSTOMER STATED THAT THE INSULIN PUMP IS ALARMING MOTOR ERROR. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290332 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LNAL

Patients

Seq Age Sex Outcome Treatment
1 74 YR