FDA Adverse Event Injury Summary report: N

SETROX S 53

MDR report key: 3812906 · Received May 15, 2014

Report

Report Number
1028232-2014-01657
Event Type
Injury
Date Received
May 15, 2014
Date of Event
March 29, 2014
Report Date
May 2, 2014
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION AND DEFORMATIONS ALONG THE LEAD BODY WHICH OCCURRED MOST LIKELY DURING THE EXPLANT PROCEDURE. DURING FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

THIS LEAD WAS EXPLANTED ONE DAY POST IMPLANT DUE TO DISLODGEMENT. THE PATIENT HAD VERY THIN VENTRICULAR WALLS, SO THIS LEAD WAS REPLACED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290361 SETROX S 53 PACER LEAD NVN BIOTRONIK SE & CO. KG 350974

Patients

Seq Age Sex Outcome Treatment
1 89 YR Hospitalization