FDA Adverse Event
Injury
Summary report: N
SETROX S 53
MDR report key: 3812906
·
Received May 15, 2014
Report
- Report Number
- 1028232-2014-01657
- Event Type
- Injury
- Date Received
- May 15, 2014
- Date of Event
- March 29, 2014
- Report Date
- May 2, 2014
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVN
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION AND DEFORMATIONS ALONG THE LEAD BODY WHICH OCCURRED MOST LIKELY DURING THE EXPLANT PROCEDURE. DURING FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.
Description of Event or Problem · 1
THIS LEAD WAS EXPLANTED ONE DAY POST IMPLANT DUE TO DISLODGEMENT. THE PATIENT HAD VERY THIN VENTRICULAR WALLS, SO THIS LEAD WAS REPLACED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290361 | SETROX S 53 | PACER LEAD | NVN | BIOTRONIK SE & CO. KG | 350974 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 89 YR | Hospitalization |