FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3812883 · Received May 15, 2014

Report

Report Number
2032227-2014-02354
Event Type
Malfunction
Date Received
May 15, 2014
Date of Event
April 23, 2014
Report Date
April 23, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH A FROZEN DISPLAY AND A CONSTANT TONE DUE TO A FAULTY INTERFACE BOARD. UNABLE TO VERIFY BLACK DOT ON THE DISPLAY AND BUTTONS UNRESPONSIVE DUE TO FROZEN DISPLAY AND CONSTANT TONE. CRACKS TO THE CASE, RESERVOIR TUBE LIP, BATTERY TUBE THREADS AND A SCRATCHED LCD WINDOW ALSO NOTED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT HE HAD LOW BLOOD GLUCOSE. CUSTOMER STATED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY, THE TIME ON THE INSULIN PUMP IS NOT ADVANCING AND THERE IS NO RESPONSE FROM ANY BUTTON. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291148 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LNAS

Patients

Seq Age Sex Outcome Treatment
1 65 YR