FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3812841 · Received April 25, 2014

Report

Report Number
3008642652-2014-01212
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
March 27, 2014
Report Date
April 17, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION MULTIPLE COMPONENTS ON THE MONITOR DEFIBRILLATOR WERE SHORTED, CAUSING THE PULSE TEST FAILURE. THE CAUSE FOR THAT DAMAGE WAS ISOLATED TO DEFECTIVE U15 AND U17 MOSFET DRIVERS ON THE DEFIBRILLATOR BOARD. THE ROOT CAUSE FOR THE DEFECTIVE U15 AND U17 COMPONENTS COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE COMPONENTS. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, MONITOR SN (B)(4) FAILED A PULSE TEST. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251337 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA