FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3812819 · Received April 25, 2014

Report

Report Number
3008642652-2014-01268
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
March 26, 2014
Report Date
April 22, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (EXPOSED WIRES) HAS BEEN CONFIRMED. UPON INVESTIGATION THE CABLE CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, DAMAGING WIRES. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT WIRES WERE EXPOSED ON HIS ELECTRODE BELT. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251177 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA