FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3812679 · Received May 15, 2014

Report

Report Number
3004209178-2014-84658
Event Type
Malfunction
Date Received
May 15, 2014
Date of Event
April 29, 2014
Report Date
April 29, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TESTING DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. A CRACKED RESERVOIR TUBE LIP, SCRATCHED LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND A CRACKED BATTERY TUBE THREADS NOTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED. THE BLOOD GLUCOSE READING WAS 193MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP WAS FLUSH. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290498 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment