FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3812671 · Received April 25, 2014

Report

Report Number
3008642652-2014-01239
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
March 20, 2014
Report Date
April 21, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK THERAPY ELECTRODE MESSAGES) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE ELECTRODE BELT FAILED A THERAPY ELECTRODE RECOGNITION TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN PULSE WIRE IN THE CABLE CONNECTING ECG "A" AND ECG "B". THERE WAS NO SIGN OF EXTERNAL DAMAGE TO THE CABLE. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE. THE PT REC'D A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS RECEIVING FREQUENT CHECK THERAPY ELECTRODE MESSAGES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251069 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 68 YR