FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3812668
·
Received April 25, 2014
Report
- Report Number
- 3008642652-2014-01235
- Event Type
- Malfunction
- Date Received
- April 25, 2014
- Date of Event
- March 24, 2014
- Report Date
- April 18, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE OR WIRES EXPOSED) WAS CONFIRMED. UPON INVESTIGATION, THE DISTRIBUTION NODE (DN) TO REAR THERAPY ELECTRODE (TE) CABLE WAS PULLED FROM THE STRAIN RELIEF. THIS RESULTED IN AN INTERMITTENT CONNECTION BETWEEN THE CABLE AND DN. THE ROOT CAUSE OF THE PULLED CABLES COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
THE MOTHER OF A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THE PT'S ELECTRODE BELT HAD EXPOSED WIRES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 250913 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |