FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 3812568 · Received April 17, 2014

Report

Report Number
3006697241-2014-00366
Event Type
Malfunction
Date Received
April 17, 2014
Date of Event
March 22, 2014
Report Date
March 22, 2014
Manufacturer
HILLROM DE MEXICO S DE RL DE CV
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND ALL 4 BRAKE CASTERS NOT HOLDING AND A BROKEN SCREW IN THE HEX ROD. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2007 AND 2008-2013. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED ALL 4 CASTERS AND THE HEX ROD TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE STRETCHER STILL ROLLS AFTER BRAKE/STEER PEDAL HAS BEEN PUT INTO THE BRAKE POSITION. THE BED WAS LOCATED AT THE ACCOUNT IN ER. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORTED WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236019 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILLROM DE MEXICO S DE RL DE CV 8050

Patients

Seq Age Sex Outcome Treatment
1