FDA Adverse Event
Malfunction
Summary report: N
CARE ASSIST BED
MDR report key: 3812564
·
Received April 17, 2014
Report
- Report Number
- 9615739-2014-00159
- Event Type
- Malfunction
- Date Received
- April 17, 2014
- Date of Event
- March 21, 2014
- Report Date
- March 21, 2014
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE CASTERS NEEDED TO BE REPLACED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012 AND 2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED ALL FOUR CASTERS AND CASTER SUPPORTS ALONG WITH THE MAIN BEARING TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES ARE NOT HOLDING. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORTED WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236001 | CARE ASSIST BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM PLUVIGNER | 1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |