Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE LEFT FOOT BRAKE CASTER WAS NOT HOLDING IN BRAKE. THE MOST LIKELY CAUSE IS DUE TO AGE AND NORMAL WEAR AND TEAR. THE TECHNICIAN SWAPPED THE RENTAL BED FOR A KNOWN WORKING BED. THE TECHNICIAN REPLACED THE LEFT FOOT BRAKE CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THAT THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED AT THE FACILITY. THERE WAS NO PATIENT/INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235914 TOTALCARE BARIATRIC PLUS BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1840

Patients

Seq Age Sex Outcome Treatment