DUAL TRIGGER ROTARY
Report
- Report Number
- 0001811755-2014-01784
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Date of Event
- April 14, 2014
- Report Date
- April 18, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THE REPORTED RUN ON WAS NOT ABLE TO BE DUPLICATED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH FUNCTIONAL EVALUATION. UPON DISASSEMBLY FOR VISUAL EXAMINATION, NO COMPONENTS WERE IDENTIFIED WHICH WOULD HAVE LIKELY CAUSED OR CONTRIBUTED TO THE REPORTED FAILURE. THE DEVICE WAS SERVICED FOR PREVENTIVE MAINTENANCE AND RETURNED TO THE USER FACILITY.
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY, THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 291823 | DUAL TRIGGER ROTARY | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | HWE | STRYKER INSTRUMENTS-KALAMAZOO |