FDA Adverse Event Injury Summary report: N

SYNCHROMED II

MDR report key: 3812329 · Received May 15, 2014

Report

Report Number
3004209178-2014-09017
Event Type
Injury
Date Received
May 15, 2014
Report Date
April 21, 2014
Manufacturer
MDT PUERTO RICO OPERATIONS CO
Product Code
LKK
PMA / PMN Number
P860004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: PRODUCT ID 8780, SERIAL # (B)(4), IMPLANTED: (B)(6) 2013, PRODUCT TYPE CATHETER. (B)(4). ANALYSIS RESULTS WERE NOT AVAILABLE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETED.

Additional Manufacturer Narrative · 1

ANALYSIS OF THE CATHETER REVEALED CATHETER BODY HAS SIGNIFICANT TWISTING THAT MAY HAVE AFFECTED INFUSION. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED UNDERDOSE SYMPTOMS, INCLUDING A SLOW LOSS OF THERAPY EFFECTIVENESS. NO SUDDEN WITHDRAWAL WAS REPORTED. THE HEALTHCARE PROVIDER (HCP) WAS UNABLE TO ASPIRATE CEREBROSPINAL FLUID (CSF) THROUGH THE CATHETER ACCESS PORT (CAP). A PARTIAL EXPLANT OF THE CATHETER WAS DONE ON (B)(6) 2014. THE CATHETER WAS TWISTED IN THE PUMP POCKET AND THE HCP WAS UNABLE TO ASPIRATE CSF. HE TRIMMED EXCESS CATHETER IN THE POCKET TO WHERE IT WAS TWISTED, REMOVED THE PUMP SEGMENT AND THEN ADDED A PUMP SEGMENT FROM THE REVISION KIT. THE SPINAL SEGMENT REMAINED INTACT. CSF FLOW WAS CONFIRMED BY THE HCP VIA THE CAP, AFTER HOOKING UP THE PUMP SEGMENT TO THE EXISTING PUMP AND AGAIN AFTER THE SUTURING OF THE PUMP TO THE POCKET. AT THE TIME OF THIS REPORT, THE PATIENT STATUS WAS INDICATED TO BE ¿ALIVE-NO INJURY.¿ THE DEVICE SYSTEM WAS USED TO DELIVER MORPHINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291768 SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MDT PUERTO RICO OPERATIONS CO 8637-20

Patients

Seq Age Sex Outcome Treatment
1 00080 YR Required Intervention