FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3812251 · Received May 15, 2014

Report

Report Number
2531779-2014-13705
Event Type
Malfunction
Date Received
May 15, 2014
Report Date
April 30, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 04/30/2014 WITH THE FOLLOWING FINDINGS: THE OK, AND UP BUTTONS RESPONDED INTERMITTENTLY TO TESTING AND THE KEYPAD COVER WAS OBSERVED TO BE PEELING AWAY FROM THE BASE. THE CONTRAST BUTTON CONTACT WAS OBSERVED TO BE MISSING AND THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED ON THE THREADS. THE DISPLAY WAS OBSERVED TO BE DIM, FADING AND REDDISH IN COLOR. THE BOLUS BUTTON COVER WAS OBSERVED TO BE TORN AND RESPONDED NORMALLY TO TESTING. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER THE UP AND DOWN BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED INTERMITTENT FUNCTIONALITY OF THE OK AND UP KEYPAD BUTTONS, A MISSING CONTRAST BUTTON CONTACT, A CRACKED BATTERY COMPARTMENT, A DIM, RED AND FADING DISPLAY SCREEN, AND CONTAMINATION UNDER THE UP AND DOWN KEYPAD BUTTON CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 04/30/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291940 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 18 YR