HOMECHOICE
Report
- Report Number
- 1416980-2014-15554
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Date of Event
- February 24, 2014
- Report Date
- April 24, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS USE ERROR, TIDAL TOTAL ULTRAFILTRATION (UF) REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 07:49:43. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 2970ML, INDICATING THE HOME PATIENT (HP) DRAINED 2570ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 291862 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |