CONTOUR NEXT
MDR report key: 3812057
·
Received May 15, 2014
Report
- Report Number
- 1826988-2014-00176
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Date of Event
- April 26, 2014
- Report Date
- April 29, 2014
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K110894
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER RECEIVED BLOOD GLUCOSE READINGS OF 386-417MG/DL ON HER CONTOUR NEXT LINK. SHE RETESTED ON ANOTHER CONTOUR NEXT LINK AND RECEIVED READINGS OF 40, 43, 38, 42, 56MG/DL. SHE THEN TESTED ON A NON BAYER METER AND THE READING WAS 160MG/DL. THE DIFFERENCES BETWEEN THE READINGS FALL IN THE "C", "D", "E" ZONES OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCES CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE CUSTOMER TEST STRIPS ARE EXPECTED TO BE RETURNED FOR EVALUATION. NEW STRIPS WERE SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 289996 | CONTOUR NEXT | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 7310 | 2JFEC32 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |