FDA Adverse Event
Malfunction
Summary report: N
CONTOUR NEXT
MDR report key: 3811997
·
Received May 15, 2014
Report
- Report Number
- 1826988-2014-00169
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Date of Event
- April 1, 2014
- Report Date
- April 23, 2014
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K110894
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INITIAL REPORTER ADDRESS AND PHONE WERE NOT PROVIDED.
Description of Event or Problem · 1
THE ADVOCATE STATED HER BROTHER RECEIVED A BLOOD GLUCOSE READING OF 190MG/DL ON THE CONTOUR NEXT LINK , WAS RETESTED ON A HOSPITAL METER AND THE READING WAS 55MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE CUSTOMER WAS ADVISED TO RETURN THE TEST STRIPS FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 292105 | CONTOUR NEXT | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 7311 | 2HFEC02 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |