FDA Adverse Event Malfunction Summary report: N

CONTOUR NEXT

MDR report key: 3811997 · Received May 15, 2014

Report

Report Number
1826988-2014-00169
Event Type
Malfunction
Date Received
May 15, 2014
Date of Event
April 1, 2014
Report Date
April 23, 2014
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K110894
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INITIAL REPORTER ADDRESS AND PHONE WERE NOT PROVIDED.

Description of Event or Problem · 1

THE ADVOCATE STATED HER BROTHER RECEIVED A BLOOD GLUCOSE READING OF 190MG/DL ON THE CONTOUR NEXT LINK , WAS RETESTED ON A HOSPITAL METER AND THE READING WAS 55MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE CUSTOMER WAS ADVISED TO RETURN THE TEST STRIPS FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
292105 CONTOUR NEXT BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE LLC 7311 2HFEC02

Patients

Seq Age Sex Outcome Treatment
1