ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-13666
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Report Date
- May 7, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
(B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT LOSS OF PRIME WARNINGS WITH NON-ZERO FORCE WAS RECORDED. DURING TESTING, THE PUMP PERFORMED THE REWIND, LOAD, AND PRIME STEPS WITH NO ALARMS OCCURRING. THE PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE WITH NO ALARMS OCCURRING. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND FADED. THE COMPLAINT THAT THE PUMP EMITTED A LOSS OF PRIME ALARM WAS OBSERVED IN THE PUMP HISTORY BUT WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. IT WAS REPORTED THAT THE PUMP EMITTED A LOSS OF PRIME WARNING AND THAT THE REPORTER HAD LOST CONFIDENCE IN THE PUMP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 291691 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR |