FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3811538 · Received May 15, 2014

Report

Report Number
2531779-2014-13562
Event Type
Malfunction
Date Received
May 15, 2014
Report Date
May 6, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/06/2014 WITH THE FOLLOWING FINDINGS: THE UP BUTTON WAS INTERMITTENTLY UNRESPONSIVE TO TESTING DUE TO A TORN KEYPAD. THE KEYPAD COVER WAS REMOVED FOR FURTHER INVESTIGATION AND THERE WAS NO CONTAMINATION FOUND UNDER ANY OF THE KEYPAD CONTACTS. UNRELATED TO THE KEYPAD ISSUE, THE PUMP BOOTED TO A DIM DISPLAY. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED INTERMITTENT FUNCTIONALITY OF THE UP KEYPAD BUTTON DUE TO A TORN KEYPAD AND A DIM DISPLAY. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 05/06/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290221 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 62 YR