ANIMAS VIBE
Report
- Report Number
- 2531779-2014-13594
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Report Date
- April 28, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 07/02/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/17/2014 WITH THE FOLLOWING FINDINGS: THERE WAS NO ILLUMINATION OF THE DISPLAY DURING STARTUP. THE PUMP DID HAVE AUDIBLE TONES DURING STARTUP. THE DISPLAY SCREEN WAS FOUND TO BE CRACKED. A TEST SCREEN WAS USED TO REPLACE THE DAMAGED SCREEN AND WORKED WITH NO ISSUES FOUND. THE DIM AND DISCOLORED ISSUE COULD NOT BE INVESTIGATED AS A RESULT OF THE BLANK DISPLAY SCREEN. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ON (B)(6) 2014, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING THAT THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290134 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |