PLM A+ SPANISH DEVIC
Report
- Report Number
- 9615050-2014-02236
- Event Type
- Malfunction
- Date Received
- March 24, 2014
- Date of Event
- January 1, 2014
- Report Date
- February 25, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K070398
- Removal / Correction Number
- FA212-05
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DEVICE DID NOT PASS THE CASSETTE TEST, THE FLUID SHIELD WAS FOUND TO BE DAMAGED, AND FLUID SPILLAGE WAS NOTED. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPITAL PERSONNEL.
DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DID NOT PASS THE CASSETTE TEST, THE FLUID SHIELD WAS DAMAGED, AND FLUID SPILLAGE WAS NOTED. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO SPECIFIC PATIENT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 172780 | PLM A+ SPANISH DEVIC | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | SN (B)(4)| PLUM A+ SOFTWARE MODULE, LIST #12097, |