PLMA DVC V11.51 1 N
Report
- Report Number
- 9615050-2014-02230
- Event Type
- Malfunction
- Date Received
- March 24, 2014
- Date of Event
- January 1, 2014
- Report Date
- March 5, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K070398
- Removal / Correction Number
- FA212-05
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AT THE USER FACILITY. DURING TESTING, THE SENSOR ON THE FLUID SHIELD WAS FOUND TO BE DAMAGED AND FLUID SPILLAGE WAS NOTED. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
DURING TESTING AT THE USER FACILITY, THE SENSOR ON THE FLUID SHIELD WAS FOUND TO BE DAMAGED AND FLUID SPILLAGE WAS NOTED. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH A NOTE THAT INDICATED MALFUNCTION N251 (CASSETTE TEST FAILED). NO TRACKING INFORMATION WAS PROVIDED; THEREFORE SPECIFIC PATIENT INFORMATION, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 172375 | PLMA DVC V11.51 1 N | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | LIST #UNK, SN UNK| PLUM + SOFTWARE MODULE, |