FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3811141 · Received March 24, 2014

Report

Report Number
1720753-2014-02624
Event Type
Malfunction
Date Received
March 24, 2014
Date of Event
March 1, 2014
Report Date
March 24, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CHOKER CONNECTING TO THE POWER SIGNAL BOARD WAS FOUND LOOS. THE CHOKER WAS RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO BOOT TO A USABLE STATE AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172559 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 9900

Patients

Seq Age Sex Outcome Treatment
1