PLUM A+ DRIVER NEW 8
Report
- Report Number
- 9615050-2014-02243
- Event Type
- Malfunction
- Date Received
- March 24, 2014
- Date of Event
- January 1, 2014
- Report Date
- February 11, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- FA210-07
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, THE DEVICE PASSED TESTING. AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE WAS NOTED IN THE DEVICE HISTORY BUT WAS NOT DUPLICATED DURING TESTING. ALTHOUGH THE DEVICE PASSED TESTING, THE DEVICE HAS BEEN IDENTIFIED A PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
DURING TESTING AT THE USER FACILITY, AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE WAS NOTED. THIS ERROR OCCURS WHEN THE BATTERY CANNOT BE FULLY CHARGED WITHIN 8 HOURS. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR A REPORT OF UNIT IS BROKEN. NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 172849 | PLUM A+ DRIVER NEW 8 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PLUM A+ SOFTWARE MODULE, LIST# UNK, SN UNK |