FDA Adverse Event
Malfunction
Summary report: N
9600
MDR report key: 3811088
·
Received March 24, 2014
Report
- Report Number
- 1720753-2014-02652
- Event Type
- Malfunction
- Date Received
- March 24, 2014
- Date of Event
- February 7, 2014
- Report Date
- March 24, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SRAM WAS EVALUATED AND IDENTIFIED AS AN END OF LIFE COMPONENT. NO REPAIR COULD BE COMPLETED AND THE CUSTOMER DISCONTINUED USE OF THE DEVICE. NO CONCLUSION CAN BE DRAWN AS CONCLUSIVE REPAIR INFO IS UNAVAILABLE AT THIS TIME AND NO ADD'L SERVICE INFO WAS PROVIDED.
Description of Event or Problem · 1
THE FSE REPORTED THAT THE SYSTEM FAILED TO ALLOW FLUOROSCOPIC EXPOSURES AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173400 | 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9600 |