FDA Adverse Event Malfunction Summary report: N

GLIDESCOPE GVL 5

MDR report key: 3810888 · Received March 19, 2014

Report

Report Number
9615393-2014-00012
Event Type
Malfunction
Date Received
March 19, 2014
Date of Event
March 18, 2014
Report Date
March 18, 2014
Manufacturer
VERATHON MEDICAL ULC.
Product Code
CCW
PMA / PMN Number
EXEMPT
Removal / Correction Number
Z-1558-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: AN INSPECTION SHOWED THE BLADE TIP WAS BROKEN OFF AT THE CAMERA. THERE WAS ALSO TIP SEPARATION, DELAMINATION, AND A CRACKED LENS COVER. THE DEVICE HAD BEEN IN USE FOR APPROX 1.5 YRS. THIS DEVICE WAS PART OF A GLIDESCOPE PRODUCT RECALL (Z-1558-2013), AND THE CUSTOMER RETURNED A CONFIRMATION FORM TO VERATHON ON (B)(4) 2013 STATING THAT USERS THROUGHOUT THEIR FACILITY WERE NOTIFIED OF THE RECALL. ON THE FORM, THE CUSTOMER WAS ASKED TO PROVIDE A LIST OF AFFECTED DEVICES IN THEIR POSSESSION, BUT THEY DID NOT LIST ANY DEVICES, AND INDICATED THAT THEY DID NOT HAVE ANY GLIDESCOPES WHICH FELL UNDER THE RECALL. IN ADDITION, A SAFETY ALERT NOTICE C/R 3022472-5-07-2013-0001-C WAS ISSUED TO ALL CUSTOMERS ON (B)(4) 2013 TO PROVIDE ADD'L SAFETY INFO TO REMIND USERS TO CAREFULLY EXAMINE BLADES BEFORE AND AFTER EACH USE, AND PROMPTLY REPLACE ANY THAT SHOW SIGNS OF WEAR OR DAMAGE. THE CUSTOMER WAS SENT A REPLACEMENT GLIDESCOPE AFTER THE COMPLAINT WAS REPORTED.

Description of Event or Problem · 1

THE PHYSICIAN STATED THAT AFTER REMOVING THE GLIDESCOPE VIDEO LARYNGOSCOPE FROM A PT, HE NOTICED A CRACK ON THE BLADE. THE CRACK WENT FROM THE TOP TO BOTTOM SEAM NEXT TO THE CAMERA. THERE WAS NO HARM TO THE PT. ON RETURN TO VERATHON, THE GLIDESCOPE TIP WAS FOUND TO BE BROKEN OFF AT THE CAMERA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163320 GLIDESCOPE GVL 5 VIDEO LARYNGOSCOPE CCW VERATHON MEDICAL ULC. 0574-0030

Patients

Seq Age Sex Outcome Treatment
1 33 YR