GLIDESCOPE GVL 5
Report
- Report Number
- 9615393-2014-00012
- Event Type
- Malfunction
- Date Received
- March 19, 2014
- Date of Event
- March 18, 2014
- Report Date
- March 18, 2014
- Manufacturer
- VERATHON MEDICAL ULC.
- Product Code
- CCW
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- Z-1558-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVAL SUMMARY: AN INSPECTION SHOWED THE BLADE TIP WAS BROKEN OFF AT THE CAMERA. THERE WAS ALSO TIP SEPARATION, DELAMINATION, AND A CRACKED LENS COVER. THE DEVICE HAD BEEN IN USE FOR APPROX 1.5 YRS. THIS DEVICE WAS PART OF A GLIDESCOPE PRODUCT RECALL (Z-1558-2013), AND THE CUSTOMER RETURNED A CONFIRMATION FORM TO VERATHON ON (B)(4) 2013 STATING THAT USERS THROUGHOUT THEIR FACILITY WERE NOTIFIED OF THE RECALL. ON THE FORM, THE CUSTOMER WAS ASKED TO PROVIDE A LIST OF AFFECTED DEVICES IN THEIR POSSESSION, BUT THEY DID NOT LIST ANY DEVICES, AND INDICATED THAT THEY DID NOT HAVE ANY GLIDESCOPES WHICH FELL UNDER THE RECALL. IN ADDITION, A SAFETY ALERT NOTICE C/R 3022472-5-07-2013-0001-C WAS ISSUED TO ALL CUSTOMERS ON (B)(4) 2013 TO PROVIDE ADD'L SAFETY INFO TO REMIND USERS TO CAREFULLY EXAMINE BLADES BEFORE AND AFTER EACH USE, AND PROMPTLY REPLACE ANY THAT SHOW SIGNS OF WEAR OR DAMAGE. THE CUSTOMER WAS SENT A REPLACEMENT GLIDESCOPE AFTER THE COMPLAINT WAS REPORTED.
THE PHYSICIAN STATED THAT AFTER REMOVING THE GLIDESCOPE VIDEO LARYNGOSCOPE FROM A PT, HE NOTICED A CRACK ON THE BLADE. THE CRACK WENT FROM THE TOP TO BOTTOM SEAM NEXT TO THE CAMERA. THERE WAS NO HARM TO THE PT. ON RETURN TO VERATHON, THE GLIDESCOPE TIP WAS FOUND TO BE BROKEN OFF AT THE CAMERA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 163320 | GLIDESCOPE GVL 5 | VIDEO LARYNGOSCOPE | CCW | VERATHON MEDICAL ULC. | 0574-0030 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR |