FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3810743
·
Received March 21, 2014
Report
- Report Number
- 8020893-2014-00673
- Event Type
- Malfunction
- Date Received
- March 21, 2014
- Report Date
- February 20, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NO INFORMATION
Narratives
Additional Manufacturer Narrative · 1
A CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB), BACKLIGHT INVERTER PCB'S AND UPGRADED THE SOFTWARE TO THE CURRENT LEVEL. THE DEVICE PASSED ALL CALIBRATIONS, SHORT SELF-TEST, EXTENDED SELF-TEST (EST) AND PERFORMANCE VERIFICATION TEST ACCORDING TO MFR SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
A REPORT RECEIVED FROM THE USA STATING THE LOWER HALF OF THE GRAPHICAL USER INTERFACE (GUI) OF AN 840 VENTILATOR IS ERRATIC. THERE WAS NO PT INVOLVEMENT REPORTED. THE DATE OF THE EVENT IS UNK. NO ADD'L INFO IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 169672 | 840 VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|