FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3810743 · Received March 21, 2014

Report

Report Number
8020893-2014-00673
Event Type
Malfunction
Date Received
March 21, 2014
Report Date
February 20, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

A CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB), BACKLIGHT INVERTER PCB'S AND UPGRADED THE SOFTWARE TO THE CURRENT LEVEL. THE DEVICE PASSED ALL CALIBRATIONS, SHORT SELF-TEST, EXTENDED SELF-TEST (EST) AND PERFORMANCE VERIFICATION TEST ACCORDING TO MFR SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

A REPORT RECEIVED FROM THE USA STATING THE LOWER HALF OF THE GRAPHICAL USER INTERFACE (GUI) OF AN 840 VENTILATOR IS ERRATIC. THERE WAS NO PT INVOLVEMENT REPORTED. THE DATE OF THE EVENT IS UNK. NO ADD'L INFO IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
169672 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1