MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2014-02999
- Event Type
- Malfunction
- Date Received
- May 14, 2014
- Date of Event
- April 15, 2014
- Report Date
- April 18, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE INSTRUMENTS REINFORCEMENT RING WAS BROKEN. THE PROXIMAL TUBE REINFORCEMENT RING WAS BROKEN. THE BROKEN FRAGMENT WAS NOT RETURNED WITH THE INSTRUMENT. FAILURE ANALYSIS CONCLUDED THAT THE BROKEN DAMAGE WAS DUE TO MISHANDLING/MISUSE. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE INSTRUMENTS MAIN TUBE WAS CRACKED. THE TUBE HAS A SMALL FISSURE IN THE AXIAL DIRECTION, LOCATED DIRECTLY BELOW THE TUBE REINFORCEMENT RING. THE LOCATION OF THE FISSURE WAS IN AN AREA THAT IS NOT PROTECTED BY THE TIP COVER. FAILURE ANALYSIS CONCLUDED THAT THE CRACKED DAMAGE WAS DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THERE WAS NO ADVERSE EVENT SINCE THE FRAGMENT WAS RETRIEVED LAPAROSCOPICALLY DURING THE SAME DA VINCI SURGICAL PROCEDURE AND NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE A FRAGMENT FROM THE MONOPOLAR CURVED SCISSORS BROKE OFF AND FELL INTO THE PATIENT.
IT WAS REPORTED THAT AFTER A DA VINCI SURGICAL PROCEDURE, THE SURGEON NOTICED A RING FROM THE MONOPOLAR CURVED SCISSORS (MCS)WAS IN THE PATIENTS ABDOMEN. THE NURSE DIDN'T NOTICE ANYTHING PARTICULAR WHEN INSTALLING THE TIP COVER ON THE MCS INSTRUMENT. WHEN THE INSTRUMENT WAS REMOVED, IT WAS OBSERVED THAT IT HAD DETERIORATED, AND THE RING WAS FOUND IN THE BODY OF THE PATIENT. ON (B)(4) 2014, INTUITIVE SURGICAL INC (ISI) CONTACTED THE INITIAL REPORTER OF THIS COMPLAINT. THE INITIAL REPORTER CONFIRMED THAT A FRAGMENT FROM THE MCS INSTRUMENT FELL INTO THE PATIENT. THE FRAGMENT WAS RETRIEVED LAPAROSCOPICALLY DURING THE SAME DA VINCI SURGICAL PROCEDURE. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED TO RETRIEVE THAT FRAGMENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. THE PATIENT HAS NOT RETURNED TO THE HOSPITAL DUE TO EXPERIENCING ANY POST SURGICAL COMPLICATIONS AS A RESULT OF THE FRAGMENT FALLING INTO THE PATIENT. THERE WAS NO ADVERSE EVENT, HARM OR INJURY TO THE INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 288556 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-12 | M10140129 483 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |