LIGASURE ADVANCE PISTOL GRIP
MDR report key: 3810456
·
Received April 15, 2014
Report
- Report Number
- 1717344-2014-00318
- Event Type
- Malfunction
- Date Received
- April 15, 2014
- Date of Event
- March 13, 2014
- Report Date
- March 17, 2014
- Manufacturer
- COVIDIEN LP
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE KNIFE IS TRAPPED AND CONTAINED WITHIN THE JAWS, WHICH KEPT THE JAWS FROM OPENING. THIS CAN OCCUR WHEN THE BLADE IS EXTENDED AND THE JAWS ARE NOT COMPLETELY CLOSED. THIS ALLOWS THE KNIFE TRACK TO OPEN TOO WIDE AND THE BLADE MOVES OUT OF ITS TRACK. THE IFU CAUTIONS TO CONFIRM THAT THE JAWS HAVE REACHED THE CLOSED POSITION AND ARE LOCKED (THE HANDLE IS LATCHED) BEFORE ACTIVATING THE CUTTER.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE HANDLE OF THE DEVICE LOCKED AFTER 1 MINUTE OF USER DURING A RIGHT SIDE HEMICOLECTOMY PROCEDURE. THERE WAS NO INJURY TO THE PT. NO FURTHER DETAILS HAVE BEEN MADE AVAILABLE BY THE SITE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230808 | LIGASURE ADVANCE PISTOL GRIP | LIGASURE VESSEL SEALING SYSTEM | GEI | COVIDIEN LP | 33430147X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |