FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® SYSTEM

MDR report key: 3810310 · Received May 14, 2014

Report

Report Number
2050012-2014-00249
Event Type
Malfunction
Date Received
May 14, 2014
Date of Event
April 14, 2014
Report Date
April 14, 2014
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE FOUND THE WASTE VALVE LEAKING FROM THE CARTRIDGE CHEMISTRY (CC) SAMPLE PROBE. THE FSE REPLACED THE VALVE AS WELL AS THE PROBE AND WASH COLLAR. PERSONAL PROTECTIVE EQUIPMENT (PPE) NUMBER WAS NOT PROVIDED. THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED SUPPRESSED (NO RESULTS GENERATED) RESULTS FOR CARTRIDGE GLUCOSE (GLU) OR MAGNESIUM (MG) ON FOUR PATIENT SAMPLES INVOLVING THE UNICEL DXC 800 SYNCHRON SYSTEM. THE CUSTOMER STATED THAT THE SAMPLES WERE REPEATED ON ANOTHER DXC ANALYZER WITH NORMAL RESULTS, WHICH WERE REPORTED. THE CUSTOMER INDICATED THAT THE QUALITY CONTROL (QC) READ IN RANGE PRIOR TO AND AFTER TESTING THE SAMPLES. THE CUSTOMER INDICATED NO ERRONEOUS RESULTS WERE GENERATED OR REPORTED ASSOCIATED WITH THIS EVENT. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288878 UNICEL® DXC 800 PRO SYNCHRON® SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1