FDA Adverse Event Malfunction Summary report: N

CENTRA BED

MDR report key: 3810294 · Received April 10, 2014

Report

Report Number
1824206-2014-01132
Event Type
Malfunction
Date Received
April 10, 2014
Date of Event
March 12, 2014
Report Date
March 12, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE BRAKE CASTER WAS OUT OF ADJUSTMENT. THE MOST LIKELY CAUSE WAS NORMAL WEAR AND TEAR. PER THE PM RECORDS, THE LAST TIME THE BRAKE CASTERS WERE REPLACED WAS 2009. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2009 THROUGH 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECH ADJUSTED THE BRAKE CASTER AND THE BED FUNCTIONED AS DESIGNED. THERE WERE NO REPORTED INJURIES. THE BED WAS LOCATED IN THE BED SHOP AT THE FACILITY.

Description of Event or Problem · 1

THE TECH WAS PERFORMING PREVENTATIVE MAINTENANCE AND REPORTED THE BRAKE WOULD NOT HOLD WHEN THE BRAKE PEDAL WAS SET TO THE BRAKE POSITION. THE BED WAS LOCATED IN THE BED SHOP AT THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
216652 CENTRA BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 850

Patients

Seq Age Sex Outcome Treatment
1