FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 3810292 · Received April 10, 2014

Report

Report Number
1824206-2014-01136
Event Type
Malfunction
Date Received
April 10, 2014
Date of Event
March 13, 2014
Report Date
March 13, 2014
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BRAKE CASTER NOT HOLDING DUE TO THE DETENT MECHANISM WORN OUT DUE TO AGE. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2006 AND 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE DETENT MECHANISM TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING BRAKES WOULD NOT HOLD. THE BED WAS LOCATED AT THE ACCOUNT IN THE HCA WAREHOUSE. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).