ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2014-13517
- Event Type
- Malfunction
- Date Received
- May 14, 2014
- Report Date
- May 7, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
(B)(4). DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS INTACT WITH NO PEELING. ALL OF THE BUTTONS WERE INTERMITTENTLY RESPONSIVE. CONTAMINATION WAS FOUND UNDER ALL OF THE BUTTON CONTACTS WHEN THE KEYPAD WAS REMOVED. THERE WAS NO EVIDENCE OF MOISTURE OR LOOSE COMPONENTS INSIDE THE PUMP WHEN THE CASE WAS REMOVED.
ON (B)(6) 2014, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING THAT THE OK AND CONTRAST KEYPAD BUTTONS WERE UNDER RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.