FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3810157 · Received February 4, 2014

Report

Report Number
3008642652-2014-00376
Event Type
Malfunction
Date Received
February 4, 2014
Date of Event
December 22, 2013
Report Date
February 3, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESPONSE BUTTONS NOT WORKING) WAS CONFIRMED. UPON INVESTIGATION THE REAR RESPONSE BUTTON WAS NON-FUNCTIONAL. THE ROOT CAUSE FOR THE NON-FUNCTIONAL RESPONSE BUTTON WAS AN INTERMITTENT CONNECTION IN THE RESPONSE BUTTON FLEX CIRCUIT. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR. THE PATIENT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) FEMALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER MONITOR RESPONSE BUTTONS WERE NOT WORKING. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
73518 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 64 YR