FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3810157
·
Received February 4, 2014
Report
- Report Number
- 3008642652-2014-00376
- Event Type
- Malfunction
- Date Received
- February 4, 2014
- Date of Event
- December 22, 2013
- Report Date
- February 3, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESPONSE BUTTONS NOT WORKING) WAS CONFIRMED. UPON INVESTIGATION THE REAR RESPONSE BUTTON WAS NON-FUNCTIONAL. THE ROOT CAUSE FOR THE NON-FUNCTIONAL RESPONSE BUTTON WAS AN INTERMITTENT CONNECTION IN THE RESPONSE BUTTON FLEX CIRCUIT. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) FEMALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER MONITOR RESPONSE BUTTONS WERE NOT WORKING. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 73518 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |