FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 3810081 · Received April 10, 2014

Report

Report Number
9615739-2014-00149
Event Type
Malfunction
Date Received
April 10, 2014
Date of Event
March 12, 2014
Report Date
March 12, 2014
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN ALLEGED HE FOUND THAT THE LEFT HEAD BRAKE CASTER AND FOOT BRAKE CASTER WOULD NOT HOLD IN BRAKE MODE. THE MOST LIKELY CAUSE WAS NORMAL WEAR AND TEAR. PER THE HILL-ROM SERVICE MANUAL THE BED SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. BRAKE CASTERS SHOULD BE CHECKED FOR CUTS, WEAR AND QUALITY OF TREAD, ETC AND REPLACED WHEN NECESSARY. CHECK THE BRAKES TO SEE WHETHER THE BED MOVES WHEN THE BRAKE PEDALS ARE PRESSED AND REPAIR AS NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2013 THROUGH 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE LEFT HEAD BRAKE CATER AND FOOT BRAKE CASTER AND BRAKE CASTER SUPPORTS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECHNICIAN FOUND THE BED WITH A NOTE THAT STATED THE BED WAS BROKEN. THE BED WAS LOCATED IN THE GROUND FLOOR HALLWAY AT THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
216270 CAREASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1