FDA Adverse Event Summary report: N

GREEN SERIES 777

MDR report key: 3809767 · Received May 2, 2014

Report

Report Number
3809767
Date Received
May 2, 2014
Date of Event
May 2, 2014
Report Date
May 2, 2014
Manufacturer
WELCH ALLYN, INC.
Product Code
DXQ
Report Source
User Facility report
Reporter Location
TX, US

Narratives

Description of Event or Problem · 1

EQUIPMENT PANEL HANGING ABOVE THE PATIENT'S BED FELL OFF BRACKET AND STRUCK THE PATIENT IN THE FOREHEAD. THE PANEL IS FASTENED TO THE WALL USING AN L-SHAPED BRACKET PROVIDED BY THE MANUFACTURER. THE BED WAS NOT AGAINST THE WALL AND THE WALL WAS NOT TOUCHED BY STAFF OR OTHER. THE BED WAS BEING MOVED INTO AN UPRIGHT POSITION BY A NURSE AT THE TIME OF THE INCIDENT. THE PANEL IN QUESTION HOLDS SEVERAL INSTRUMENTS, INCLUDING AN OTOSCOPE, A THERMOMETER, AND BLOOD PRESSURE CUFF. THE PATIENT WAS TAKEN FOR A CT SCAN, WITH NEGATIVE RESULTS OTHER THAN MINOR SOFT-TISSUE DAMAGE. THE BIOMED STAFF HAS CONCERNS WITH THE WAY THE PANEL IS DESIGNED IN THE BACK IN THAT IT COULD BE INADVERTENTLY KNOCKED OFF THE WALL AS IT IS ONLY HELD BY GRAVITY. OUR 'WORK AROUND' HAS BEEN TO SCREW IN THE CENTER OF THE BOTTOM OF THE PANEL TO THE WALL, SO IT CANNOT BE LIFTED UP. WE HAVE MADE THE MANUFACTURER AWARE OF THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
265506 GREEN SERIES 777 INTEGRATED WALL PANEL FOR ANEROID MANOMETER, OTOSCOPE, THERM DXQ WELCH ALLYN, INC. * *

Patients

Seq Age Sex Outcome Treatment
1 86 YR