FDA Adverse Event Malfunction Summary report: N

PROTEGE CEILING MOUNT MICROSCOPE SYSTEM

MDR report key: 380732 · Received February 28, 2002

Report

Report Number
1937051-2002-00001
Event Type
Malfunction
Date Received
February 28, 2002
Date of Event
January 7, 2002
Report Date
February 28, 2002
Manufacturer
GLOBAL SURGICAL CORP.
Product Code
HRM
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CEILING MOUNTED MICROSCOPE WAS BEING MOVED INTO ITS STORAGE POSITION WHEN THE MICROSCOPE FELL TO THE FLOOR. THERE WERE NO INJURIES OR PTS INVOLVED. THE MOUNTING POST ASSEMBLY THAT SECURED THE MICROSCOPE ASSEMBLY TO THE CEILING WAS RETURNED AND EVALUATED BY THE MFR. EXAMINATION OF THE 2 SETSCREW MARKS ON THE SHAFT INDICATES THAT THE POST WAS NOT FULLY INSTALLED INTO ITS MOUNT DURING THE ORIGINAL ASSEMBLY. SECONDARY SETSCREW MARKS ON THE SHAFT INDICATE THAT THE SETSCREWS WERE RE-TIGHTENED AT SOME POINT, WITH THE SHAFT UN-THREADED BY ABOUT ONE ROTATION. THE DESIGN OF THIS DEVICE WAS CHANGED IN MAY OF 1999. THE ASSEMBLY IS NOW DRILLED AND PINNED, THUS REPLACING THE SET SCREWS WITH 2 DOWEL PINS THAT PREVENT THE ASSEMBLY FROM LOOSENING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROTEGE CEILING MOUNT MICROSCOPE SYSTEM DENTAL MICROSCOPE HRM GLOBAL SURGICAL CORP. M 705C NA

Patients

Seq Age Sex Outcome Treatment
1 NA